Cameras & sensorsExplainer

Camera sensor technologies by spectral range: CMOS to microbolometer

Match detector physics to band: silicon CMOS 400–1000 nm, UV-enhanced BSI below 400 nm, InGaAs 0.9–2 µm, cooled MWIR assemblies, microbolometer LWIR 8–14 µm.

LumaSpec Vision10 min read Share on LinkedIn

On this page 9 sections
  1. 01Which sensor technology covers each wavelength band?
  2. 02What can a standard silicon CMOS camera actually see?
  3. 03When does UV imaging need more than a standard sensor?
  4. 04Why does SWIR need InGaAs instead of silicon?
  5. 05Why is MWIR a cooled detector assembly, not a camera sensor?
  6. 06How does an uncooled microbolometer measure heat at 8–14 µm?
  7. 07Can a thermal camera resolve your defect?
  8. 08Which sensor technology should you choose?
  9. 09Questions

A camera sees only the wavelengths its detector physics absorbs, so the spectral range is a sensor decision before it is a software one. Pick the wrong family and the defect you are chasing is genuinely invisible: no lens, light or algorithm recovers it. This guide maps the five detector families used in industrial imaging — silicon CMOS, UV-enhanced backside-illuminated CMOS, InGaAs, cooled MWIR assemblies and uncooled microbolometers — to the bands they cover, the cooling and optics each one forces on you, and a pixel-budget example that shows when a thermal camera cannot resolve a defect at all. It is written for the engineer who has a signal or a defect in mind and needs the simplest detector that still sees it.

Which sensor technology covers each wavelength band?

Each band of the spectrum is served by a different detector: silicon CMOS for 400–1000 nm, UV-enhanced backside-illuminated CMOS below 400 nm, InGaAs for 0.9–2 µm, cooled InSb or XBn assemblies for 3–5 µm, and uncooled microbolometers for 8–14 µm. Two boundaries drive the whole table. Silicon stops absorbing near 1.1 µm, which sets the long edge for CMOS and everything built on it. Below roughly 400 nm, ordinary glass and sensor cover glasses absorb, so UV work needs quartz optics and UV-capable sensors.

Band Detector physics Typical industrial jobs
UVA 315–400 nm Standard BSI mono CMOS Fluorescence, adhesive cure checks
UVB 280–315 nm, UVC below 280 nm UV-enhanced BSI CMOS Short-wave UV imaging, contamination
VIS 400–700 nm Silicon CMOS, mono or colour Code reading, measurement, most inspection
VNIR 700–1000 nm Silicon CMOS mono 850–920 nm LED illumination, hyperspectral
SWIR 0.9–2 µm InGaAs on a CMOS readout Moisture, wafers, plastic sorting
MWIR 3–5 µm Cooled InSb or XBn-InAsSb Gas imaging, black plastic, active thermography
LWIR 8–14 µm Uncooled microbolometer Thermography, process monitoring

X-ray imaging sits outside the table as a scintillator in front of a silicon sensor: the crystal converts X-ray photons to visible light and the CMOS chip reads the result. Everything else in an industrial camera is one of the rows above, and the area scan camera and infrared camera ranges split along exactly these lines.

What can a standard silicon CMOS camera actually see?

A mono silicon CMOS sensor covers 400–1000 nm without modification, so it sees the visible spectrum plus the 850–920 nm band where common IR LEDs emit; colour versions trade that near-IR response for a Bayer mosaic and a debayered RGB image. That single span carries the majority of machine vision: code reading, measurement, presence checks, pick and place. Inside it you choose three things. Mono versus colour, where mono keeps every pixel and the full near-IR response. Interface, GigE for cable length, USB3 for simplicity. And the lens, which must resolve the pixel pitch — matching sensor to lens at the Nyquist limit is that pairing check.

Area scan cameras12 MP USB3 area scan camera, IMX226, 1/1.7" rolling shutter, 30 fps, 4000 × 3000 resolution (LS-CMR-0166)LS-CMR-0166
Sensor format
1/1.7"
Interface
USB3
Resolution
12 MP
Shutter
Rolling
$560.00View partBuild a system with it

The LS-CMR-0166 is the archetypal VIS/VNIR workhorse: 12 MP of silicon in 1/1.7" format, 4000 × 3000 pixels read out at 30 fps over USB3. For most plants, this row is the whole answer, and the families below exist only for the signals silicon cannot absorb.

When does UV imaging need more than a standard sensor?

Standard glass and standard silicon both fade below about 400 nm, so UVA work at 315–400 nm runs on BSI mono sensors, while UVB at 280–315 nm and UVC below 280 nm need UV-enhanced sensors engineered for short-wave quantum efficiency. Backside illumination is the enabling change: flipping the sensor puts the photodiode in front of the transistor stack, so UV photons are absorbed before the metal layers eat them. On a BSI mono camera, UVA fluorescence is now an ordinary industrial build. Below 315 nm, quantum efficiency falls away in a conventional silicon stack, and UV-enhanced sensor designs are the practical route. The optics change with the band: glass absorbs short-wave UV, so lenses go quartz or UV-corrected — what changes in a lens outside visible light covers the materials.

Why does SWIR need InGaAs instead of silicon?

Silicon stops absorbing near 1.1 µm, so imaging 0.9–2 µm needs indium gallium arsenide, a hybrid detector in which an InGaAs photodiode layer is bump-bonded to a silicon readout circuit similar to a global-shutter CMOS sensor. Off-the-shelf industrial InGaAs cameras now reach around 5 MP, but three costs come with the band. Pixel pitch runs 10–20 µm against 2–5 µm for modern silicon, so the same format carries far fewer pixels. Dark current is much higher, so the detector is thermoelectrically cooled tens of degrees below ambient, and long-exposure scientific builds go far deeper. And InGaAs detectors cost several times their silicon equivalents.

Infrared camerasShort-wave infrared area scan camera, 0.3 MP, GigE, Mono, 15μm x 15μm pixels (LS-IRC-0029)LS-IRC-0029
Sensor format
1/1.2"
Lens mount
C-mount
$5,999.00 · quotedView and quote
Infrared camerasHikrobot MV-CI013-GS-NN 1.3 MP SWIR Camera, IMX990, GigE, fanlessMV-CI013-GS-NN
Sensor format
1/2"
Lens mount
C-mount
Price on requestView and quote

Both cards are GigE industrial SWIR cameras: the LS-IRC-0029 with 0.3 MP on 15 µm × 15 µm pixels, the fanless MV-CI013-GS-NN at 1.3 MP on the IMX990 sensor. Choose InGaAs when the contrast only exists up there — moisture absorption, wafers under opaque films, polymer sorting past 1.1 µm — starting from which band to inspect with: SWIR, NIR or visible. Long-exposure builds add drift questions, covered in uncooled SWIR camera stability.

Why is MWIR a cooled detector assembly, not a camera sensor?

At 3–5 µm the dark current of suitable semiconductors is so high that the detector must be chilled: InSb runs near 77 K (−196 °C) behind an integral cryocooler, XBn-InAsSb works around 150 K, about −123 °C, so you buy a sealed assembly, not a board-level sensor. The 3.6–4.2 µm band is where XBn barrier detectors earn their complexity, and InSb stays the reference for high-speed thermal events. What justifies the cooler: gas imaging, since hydrocarbon bonds absorb strongly near 3.3 µm and a leak reads as a hot plume; black plastics, which carbon-black loading makes indistinguishable in NIR sorting but separable in MWIR; and active thermography on metals, where heat diffuses fast enough that only a high-frame-rate cooled imager catches the crack.

Infrared camerasGas leak imaging camera, GigE, 50 mm lens (LS-IRC-0007)LS-IRC-0007
Interface
GigE
Spectrum
MWIR
$47,000.00 · quotedView and quote

MWIR is bought as an engineered detector unit with its own cryocooler and control electronics, so the models are quoted items rather than shelf cameras — plan the lead time and the mounting heat load into the project.

How does an uncooled microbolometer measure heat at 8–14 µm?

A microbolometer is a thermal detector: LWIR radiation at 8–14 µm warms a membrane whose resistance changes, giving 30–40 mK NETD at 30–60 Hz without any cryogenic cooling, which is why every industrial thermography camera uses one. Because the readout is thermal, not photonic, the trade-offs are temporal and spatial: response is relatively slow, arrays run QVGA to VGA with megapixel parts appearing, and the germanium lens is calibrated to the sensor as one unit, so optics are not swappable afterwards. The common temperature span for industrial radiometry is about -10 to 140 °C, and 9 Hz variants exist chiefly to stay under export-control thresholds.

Infrared cameras640 x 512 thermal camera, 8-14 μm, 50 Hz, GigE, FOV 88.5° × 73.2°, 9-24 V DC (LS-IRC-0044)LS-IRC-0044
Lens mount
C-mount
Interface
GigE
Resolution
Up to 3 MP
Spectrum
LWIR
$5,885.00 · quotedView and quote

The LS-IRC-0044 is the class in one card: 640 × 512 pixels on an 8–14 µm band, streamed at 50 Hz over GigE through an 88.5° × 73.2° lens. For monitoring work, NETD and lens choice for process monitoring sets the deciding numbers; for measurement, raw thermal counts to temperature is the calibration step after the optics.

Can a thermal camera resolve your defect?

Often no: budget three pixels across the smallest feature, divide field of view by pixel scale, and a 640 × 512 microbolometer on a 100 mm field reaches only 0.16 mm/px, so any feature under about 0.5 mm is gone regardless of its 30–40 mK sensitivity. The arithmetic is two lines: pixel scale equals field of view divided by pixels across the sensor, and a feature needs about three pixels to be detected reliably.

pixel scale = field of view ÷ pixels across feature size needed = pixel scale × 3

Worked example: a 0.25 mm crack on a 100 mm wide machined face.

  • Required scale: 0.25 mm ÷ 3 = 0.083 mm/px.
  • Required array: 100 mm ÷ 0.083 mm/px ≈ 1200 px across.
  • Silicon at 4000 px: 100 mm ÷ 4000 = 0.025 mm/px, so the crack spans 0.25 ÷ 0.025 = 10 px. Detected and measurable.
  • Microbolometer at 640 px: 100 mm ÷ 640 ≈ 0.16 mm/px, so the crack spans 0.25 ÷ 0.16 ≈ 1.6 px. Not detectable, whatever the NETD table says.

The arithmetic lands on InGaAs too: at 15 µm pitch an InGaAs sensor carries about a third of the pixels per line of a 5 µm silicon part in the same format, so the SWIR field of view shrinks or the magnification drops. Run your own numbers in the camera resolution calculator.

Camera resolution for the smallest feature
Units
Pixels needed
–
Sensor size
– MP
Smallest sensor that fits
–
Pixel on the part
– mm
Pixels across the feature, that sensor
–

Pixels needed = field / feature × pixels per feature on each axis (field incl. margin). The lens must resolve the pixels too: Nyquist = 1000 / (2 × pixel µm) lp/mm.

Find cameras with enough pixels Open the full calculator

Which sensor technology should you choose?

Choose the least exotic detector that still sees the defect: silicon for 400–1000 nm, UV-enhanced BSI below 400 nm, InGaAs from 0.9 µm, cooled assemblies only when 3–5 µm absorption or fast thermal events demand it, and microbolometers whenever the signal is surface temperature. Routing by case:

  • Code, geometry, colour, surface: silicon CMOS, mono unless you truly need colour — the camera finder filters by sensor, resolution and interface, and the camera step of the configurator builds the chain around your pick.
  • UVA fluorescence, 315–400 nm: BSI mono camera with quartz optics.
  • UVB and UVC below 315 nm: UV-enhanced sensor, quartz optics, interlocked illuminator.
  • Contrast only above 1.1 µm, moisture, wafers, polymer sorting: InGaAs SWIR.
  • Gas, black plastic, fast thermal events: cooled MWIR assembly.
  • Surface temperature from about -10 to 140 °C, hot spots, delamination by cooling behaviour: microbolometer LWIR.
  • Material identity rather than presence: hyperspectral cameras splitting the silicon band itself — the LS-HSC-0004 resolves 400–1000 nm into 1200 bands over USB3 — with how hyperspectral imaging works as the method check.
Hyperspectral cameras400-1000 nm area scan hyperspectral camera, 1200 bands, USB3 (LS-HSC-0004)LS-HSC-0004
Lens mount
C-mount
Interface
USB3
Resolution
5 MP
$15,900.00 · quotedView and quote

Two checks kill more spectral builds than sensor choice does. Lens transmission: glass stops near 1.1 µm on the long side and below about 350 nm on the short side, so quartz or germanium optics are part of the order, not an accessory. Cooling duty: a thermoelectrically cooled SWIR camera or a cryogenically cooled MWIR unit must reject its own heat, which decides the mounting before the first image. If two rows still look plausible, re-run the pixel budget above, or send the part and the defect to expert support — we will match band, array and interface in one pass. The cameras topic page collects the deeper guides on each family.

Questions

Why does a colour camera miss my 850 nm infrared illumination?

No — and this trips more builds than sensor physics does. Mono silicon CMOS keeps useful response to about 1000 nm, so 850–920 nm LED illumination works on any standard mono industrial camera. A colour model is the wrong base: its Bayer mosaic and filter stack discard most of the near-IR signal, so build IR applications monochrome, with IR-transmitting optics.

Do InGaAs SWIR cameras need cooling?

Yes, at least thermoelectric cooling. InGaAs dark current is orders of magnitude higher than silicon dark current at room temperature, so industrial SWIR cameras hold the detector tens of degrees below ambient, and deep-cooled scientific models go far lower for long exposures. Bright, short exposures tolerate less cooling, but dark current still sets the noise floor of the image.

Can I fit a normal C-mount glass lens to a thermal camera?

Glass does not transmit 10 µm radiation, so thermal cameras use germanium optics calibrated to that specific sensor as part of the radiometric calibration. Swapping the lens shifts the temperature readings and usually breaks the calibration. Choose the field of view at order time; the QVGA or VGA array and its germanium lens ship matched.

What NETD do I need for industrial thermography?

For process monitoring, the standard 30–40 mK is enough: it is the smallest temperature step the camera separates from noise, and it tells you whether a hot spot or a cooling anomaly is real. Subtle delamination work by active thermography wants lower NETD, and metals that conduct heat quickly push you to a fast cooled MWIR imager instead.

Is a 9 Hz thermal camera enough for industrial inspection?

For static or slowly changing surfaces, yes — 9 Hz variants exist to stay under export-control thresholds for dual-use goods, not because the detector is weaker. The moment the job involves active thermography, flash excitation, metal surfaces or anything at production speed, you need the 30–60 Hz class, and export rules then become part of the project plan.